6 Logging
19
Computer as output medium
The following conditions must be fulfilled in order to use the computer as an output medium:
ü
The computer is equipped with an USB port.
ü
The MELAG USB driver is installed on the computer.
ü
The sealing device is connected to the computer via a USB cable (optionally available).
ü
Firmware 1.12 requires MELAview Doku/MELAtrace version 3.0.15.1.
ü
Firmware 2.10 and above requires MELAview Doku/MELAtrace from version 3.6.2.1.
1.
Install the MELAseal 200 PC driver on your computer.
The driver can be downloaded from the MELAG
website (Service/Download Center).
2.
Connect the type B USB cable to the rear panel of
the sealing device and connect the other end to the
computer.
3.
Establish a connection between the documentation
software and the sealing device and determine the
folder in which the sealing logs are to be archived.
Further information relating to the operation of the
sealing device with the documentation software can
be found in the corresponding operating manual.
PLEASE NOTE
The daily log is continued even if the sealing device is switched on and off once; i.e. a
new daily log is not started.
Structure of the log files
The log is stored in HTML format (in English). The file name consists of the actual date, the serial number
of the sealing device and the total counter. A log includes all printing procedures of one day. A row in the
log is written for every sealing procedure.
20130723_132000901_0000095.htm
Date
Serial number
Total counter
A successful sealing procedure is marked with “OK” in the respective log row. If a malfunction occurs dur-
ing a sealing procedure, this is recorded in the log by a incident number, e.g. F4. The corresponding fault
value (e.g. sealing temperature, or force) will, depending on the nature of the malfunction, be displayed in
square brackets or no value will be displayed.