The Cal Kit panel shows the properties of the selected Cal Kit in the list. The name and
label can be changed. The label is used for a short notation of the kit in an element
definition.
The Cal Kit Elements panel shows the calibration elements in the selected Cal Kit. You
can again add new elements, export and import them from a Cal Element file (.vce).
The bottom panel shows the properties of the selected Cal Element. On the left are general
properties and on the right are the coefficients or S-parameters that define the
measurement compensation for the element.
The
Label
is used for short notation of the calibration element. The
frequency range
is
used to determine if an element is suited for the current measurement range. A zero value
for the max frequency causes the software to ignore the max frequency.
The type selection identifies the
type of calibration
and the connector and gender boxes
allow for the selection of
connector
and
gender
of the element. The
serial number
allows
identifying a specific Cal Element from a specific Kit.
There are actually three Genders:
Female
,
Male
and
Neutral
. Neutral can be used for
sexless connectors or for test fixtures and PCB Cal Kits that don’t calibrate at a connector.
Neutral is
not compatible
with Female or Male.
The
Gender
is that of the
Cal Kit Element itself
while the gender of the port setup is that
of the measurement port. The software will
match a Male Element with a Female Port
and Neutral with Neutral
.
Cal Element definitions
The electrical properties of the calibration elements can be defined with
Coefficients
or
by measured
S-parameters
.
In the
Coefficient
mode all element have a
Delay
and
Loss
property. Some elements have
a small length of transmission line (or adapter). The Delay is the distance (travel time)
from the reference plane to the actual calibration impedance. The Loss is the skin effect
loss of that transmission line.
The other coefficient definitions depend of the type of the Cal Element.
Open Element