Single-port calibration (OSL)
If you are performing a 1-port measurement, you only need the Open, Short and Load.
Instrument set-up
•
Select the kind of measurement (1-port), frequency range, generator power, etc.
•
Select the
Calibration
tab. This shows the calibrations that are necessary for the
measurement as well as the configuration of the measurement port(s).
•
Set up the port configuration with the
Port Setup
button.
•
In the Port Configuration form, select the
Connector
type and
Gender
for the VNA
Sandbox. The program will search for the correct Calibration Kit elements. Press
OK
.
•
Back in the calibration tab, the calibrations in green are already completed and the
ones in red still need to be done. With the
Calibrate
button you can step through the
calibrations. But the calibrations can also be made in any other order.