84
9.
After selecting the appropriate tap, select the play button and a connection diagram will appear
showing how to correctly connect the MRCT to the CT
Figure 58: CT Winding Resistance Connection Diagram Non Concurrent
10.
Check all connections to ensure correctness
11.
Hit Blue Play button to begin test.
12.
The Winding Resistance test will begin.
X Voltage (V): The MRCT displays the real time voltage readings of the secondary X winding
while the test is in progress.
X Current (A): The MRCT the real time DC current readings for the secondary X winding while the
test is in progress.
13.
At any point of time Test can be terminated by hitting the ABORT button at the top