1. Switch the instrument ON
2. Ensure the mains supply to the charger is powered up and
turned on
3. Plug the EVCC300 into the charger outlet, either directly into
a type 2 connector or using the adaptor provided into a type
1 connector
Ensure the connection method indicated in the top left corner of
the MENU screen is correct.
As a reminder the menu screen connection indicator looks like
this.
In this screen the EVCC300 is set to direct connection. If the red
cross is not present the instrument is set for connection to the
charger via a tethered cable or plug in charging cable.
4. Use softkeys 1 and 3 (LEFT and RIGHT)to select the
“Interface”test indicated in move
5. Press softkey 2 to SELECT the test
6. The EVSE Interface screen will appear
The measurements on the EVSE interface test are single
measurements and are not updated continuously. After a short
time warning triangles will appear next to each measurement to
indicate that these measurements are held measurements and
are not being updated. To update the measurements press and
hold for short period the TEST button.
7. Press softkeys 1and 2 to select the desired CP code applied
to the charger. At this point no measurements are taken,
only the CP code applied to the charger.