11. Specifications
Specification
Detail
PHYSICAL
Dimensions:
228 x 105 x 75 mm (8.98 x 4.1 x 2.95 in)
Display:
Full LCD colour screen with user configurable backlight.
Weight:
905 grams (with batteries)
SAFETY AND ELECTRICAL PROTECTION
Safety rating
CATIII 600 V / CATIV 300 V to EN 61010, IEC 61010-031 : 2015, IEC 61010-030
Safety category rating valid to altitude of 3000 m
Live voltage
Active live voltage protection to 600 V between any test terminals without blowing a fuse.
Live voltage warning on display and audible when >5 V is applied between any test terminals.
Fuse protected to 1000 V, fuses are not user changeable.
LOW RESISTANCE MEASUREMENT
Resistance measurement test modes:
Normal test mode, Fast mΩ/long test lead mode and Inductive test mode (Resistance of
inductive loads)
Overall resistance range:
1
μΩ
– 2000 Ω
Max resistance across C terminals:
2 A with up to 1.1 Ω total resistance and 1 A with up to 3.2 Ω total resistance
Basic accuracy:
Bi-directional test current mode: + /- 1% +/- 2 digits
Uni-directional test current mode: + /- 1% +/- 10 digits
Inductive mode or unidirectional mode will introduce an undefined error
if an external EMF is present
Resistance measurement ranges:
Full Scale
Resistance
Test Current
Resolution
Normal
Resistance
Mode
Inductive
Mode
Long Test
Lead Mode
(1 A only)
15000 μΩ
2.00 A
1 μΩ
120.00 mΩ
2.00 A
0.01 mΩ
1000.0 mΩ
2.00 A
0.1 mΩ
30.000 mΩ
1.00 A
0.001 mΩ
220.00 mΩ
1.00 A
0.01 mΩ
2000.0 mΩ
1.00 A
0.1 mΩ
300.00 mΩ
100 mA
0.01 mΩ
2500.0 mΩ
100 mA
0.1 mΩ
20.000 Ω
100 mA
0.001 Ω
3000.0 mΩ
10.0 mA
0.1 mΩ
24.000 Ω
10.0 mA
0.001 Ω
200.00 Ω
10.0 mA
0.01 Ω
30.000 Ω
1.00 mA
0.001 Ω
240.00 Ω
1.00 mA
0.01 Ω
2000.0 Ω
1.00 mA
0.1 Ω
Specifications