OPERATION

7

OPERATION

the settings should be 9600 baud, 8 data bits, 1 stop bit, and no parity.

Downloading data does not cause the stored data to be erased from 
memory. To clear data from memory see “

Delete Data

” below.

Please note – 

 

 DLRO600 also makes data available via the RS232 

Port in real time and is suitable for printing on a self-powered serial 
printer. (See later for details.)

The output has the following form:

 

Test Type

 

Test Number

 

Date Format 

 

 Date

 Time

 

Selected Current 

 Resistance

 

Measured Current

 

Measured Voltage

 

Upper Limit  *

 

Lower Limit  *

 

Pass Or Fail  *

* The last three lines will only appear if passbands have been set.

measure will then activate the measurement.

AUTOMATIC

 mode is abbreviated to 

AUTO

 under the 

TEST

 heading in 

the main menu screen. 

AUTO

 mode may be exited by pressing the 

TEST

 

button while the status light is flashing.

OPTIONS Menu

The Options menu has five selections, which control various ancillary 
functions and which are selected using the two-axis cursor control and  
the Enter key. These are Retrieve, Passbands, Set Clock, Delete Data and 
Storage.

Retrieve

allows the recall of results which have been stored within the DLRO600’s 
internal memory. There are two choices, Display or Download.

Display

recalls each test, in sequence, to the instrument display starting with the 
most recently stored result. Use the cursor control Up and Down controls 
to step later or earlier respectively through the stored results.

If you know the number of the test that you wish to display simply type in 
the number and press Enter.

An asterisk (

*

) next to the word “

MEMO

” at the bottom of the screen 

means that there are notes attached to this result. Press the right cursor 
control to view the notes.

Download

causes the entire contents of the data store to be output to the RS232 
port above the display. A null modem RS232 lead is supplied with the 
instrument. Any communications package may be used to download the 
data although a copy of Download Manager, which facilitates downloading 
and formats the data, is also supplied. If using another package please note 

Содержание DLRO 600

Страница 1: ...M DLRO 600 High Current Low Resistance Ohmmeter USER MANUAL...

Страница 2: ...le generator where the two supply wires are floating Connecting the DLRO ground to real ground makes it safe but the internal test circuit does not work because there is no connection between ground a...

Страница 3: ...n the normal waste stream CONTENTS 3 SAFETY WARNINGS 2 GENERAL DESCRIPTION 4 OPERATION 5 Main menu screen 5 TEST menu 6 OPTIONS menu 7 CURRENT menu 9 Warning messages 9 Entering notes in the memo scre...

Страница 4: ...tance of the test leads providing sufficient smoothing for the current P1 and P2 are voltage measuring terminals P1 is nominally positive but connections can be reversed without affecting the measurem...

Страница 5: ...ypad When finished press Enter If the current you have set is greater than the current limit DLRO600 will beep and clear the SET CURRENT field Re enter the desired current if necessary increasing the...

Страница 6: ...age leads to be securely connected to the item under test before pressing the Test button The DLRO600 checks voltage and current loop continuity as in Normal Mode and if adequate passes a continuous d...

Страница 7: ...which are selected using the two axis cursor control and the Enter key These are Retrieve Passbands Set Clock Delete Data and Storage Retrieve allows the recall of results which have been stored with...

Страница 8: ...main menu screen will also display 300 MEMORY FULL instead of the number of stored tests Passbands The passband option allows you to set upper and lower limits between which the test result must lie i...

Страница 9: ...uit a message will appear on the display stating POTENTIAL FAIL Press Enter to clear this message and correct the discontinuity in the P circuit Press TEST again If the P circuit is complete DLRO600 w...

Страница 10: ...f the output is being monitored when the DLRO600 is switched on you will capture the instrument type DLRO600 and the version of firmware being run in the instrument Thereafter the data will be TEST NU...

Страница 11: ...leads should be adequately insulated and sufficiently strong mechanically to support their own weight if long leads are being used The size of the current leads will limit the maximum resistance which...

Страница 12: ...x 250 x 250 mm Weight 15 kg Measurement Ranges Full Scale Resolution Volts full scale Max Test Current 999 9 0 1 0 6 V 600 A 9 999 m 1 3 V 300 A 99 99 m 10 4 V 40 A 999 9 m 100 4 V 4 A Subject to supp...

Страница 13: ...batteries in other parts of the EU contact your local distributor Megger is registered in the UK as a producer of batteries The registration number is BPRN00142 REPAIR AND WARRANTY 13 The instrument...

Страница 14: ...14...

Страница 15: ...14 337 3038 Megger Z A Du Buisson de la Couldre 23 rue Eug ne Henaff 78190 TRAPPES France T 33 1 30 16 08 90 F 33 1 34 61 23 77 Megger products are distributed in 146 countries worldwide This instrume...

Отзывы: