M
DLRO10HD
10 A Digital Low Resistance Ohmmeter
USER GUIDE
Страница 1: ...M DLRO10HD 10 A Digital Low Resistance Ohmmeter USER GUIDE...
Страница 2: ...the LCD will flash at end of a test on an inductive load whilst a discharge current is still flowing and will stop flashing when the current has decayed to less than 1 mA For testing large inductive r...
Страница 3: ...onnectors to confirm that their condition is good with no damaged or broken insulation Charging and mains line operation In normal circumstances the instrument is completely safe However in case of an...
Страница 4: ...the instrument and wipe it with a clean cloth slightly damped with soapy water or Isopropyl alcohol IPA Symbols used on the instrument G Caution Refer to accompanying notes F HV warning t Equipment pr...
Страница 5: ...10 Amp tests to be completed on a single charge An auto power off function preserves battery life when operating from battery power The battery contains intelligent circuitry which prevents damage to...
Страница 6: ...achieved Electromagnetic noise induced into the leads can interfere with a reading A noise symbol alerts the user and prevents a measurement when the instrument detects noise above its threshold When...
Страница 7: ...connect the test leads and press the test button The measurement is updated every three seconds until the circuit is broken Please note that in this mode both the current and voltage leads must be co...
Страница 8: ...rnal voltage is applied between the terminals and the instrument is on the HV warning Fwill flash on the display This is a warning that the item under test is live and might be dangerous and testing i...
Страница 9: ...ontacts connect L1 will extinguish 4 No lamps will show during the test unless contact fails 5 Lamp L2 will light a continuous green when current flow has decayed to less that 1 mA to signal end of te...
Страница 10: ...the contrast adjust button Contrast setting will scroll through its range from min to max and back to min etc The backlight ON OFF button toggles the state of backlight that will timeout if left on f...
Страница 11: ...th arrows to indicate direction of current flow The large arrow at the top of the display between C1 and C2 indicates measurement current flow LCD Icons Battery icons with battery charge level indicat...
Страница 12: ...fitted with a 3 Amp fuse The instrument can be powered from 90 265 Vrms a c at 50 60 Hz The battery will charge as long as the mains supply is connected except when a test is in progress For optimum b...
Страница 13: ...nal mode will introduce an undefined error if an external EMF is present Basic accuracy at reference conditions Test Current Resistance Range Resolution as displayed Basic Accuracy Full scale voltage...
Страница 14: ...ode selected IP rating IP65 case closed IP54 battery operation Safety rating In accordance with IEC61010 1 CATIII 300 V when used with optional terminal cover details in ordering information Applicati...
Страница 15: ...s at the instrument terminals Although the terminal cover may be used with any test leads only Megger DH4 and DH5 duplex handspikes have suitable probe insulation to comply with the requirements of IE...
Страница 16: ...the printed circuit board If the protection of an instrument has been impaired it should not be used and be sent for repair by suitably trained and qualified personnel The protection is likely to be...
Страница 17: ...oducts is a reminder not to dispose of the product at the end of it s product life with general waste Megger is registered in the UK as a Producer of Electrical and Electronic Equipment The Registrati...
Страница 18: ...pliance with Directive 2014 53 EU Other equipment manufactured by Megger Instruments Limited described in this user guide is in compliance with Directives 2014 30 EU and 2014 35 EU where they apply Th...
Страница 19: ...enaff 78190 TRAPPES France T 33 0 1 30 16 08 90 F 33 0 1 34 61 23 77 E infos megger com Other technical sales offices Toronto Canada Sydney Australia Mumbai India Madrid Spain and the Kingdom of Bahra...