International model variations:
Continuity retest after fail (PAT120, PAT150 DE, & CH models only)
When a continuity test fails to meet the pre-set continuity resistance pass limit of 0.3
W
, the test can be run again within 5 seconds at the higher 1.0
W
limit.
Example Class I continuity FAIL. Display shows:
to retest at 1.0
W
limit or
to FAIL test