20
3.1.2 Charge the PDM
As a
safety measure, the PDM will give you “PDM BATTERY LOW” or “CHARGE
PDM NOW” alert when you keep the PDM working at a low power level. If you
receive a “PDM BATTERY LOW” alert, respond to the alert and continue. Though
the PDM will still function normally, the battery life could be decreased.
The PDM requires an AC adapter with an output of DC 5.0V that complies with
IEC 60601-1 and IEC 60950 such as UES06WNCPU-050 100SPA, (input: 100-240V,
50/60Hz, 0.2A; output: 5.0V DC, 1.0A). The adapter is designed as a part of the
ME system.
Note:
⚫
Do not use other types of chargers. Otherwise the PDM may not work
normally.
⚫
You must charge the PDM when the battery is low to keep using the PDM.
If the battery is exhausted, the PDM will shut down automatically.
⚫
No settings will be lost if the PDM power is depleted or PDM error happens.
⚫
The battery must be fully charged the first time you use the PDM, which
usually takes around 2 hours. If the battery is not full after 12 hours of
continuous charging as required, please contact customer service.
⚫
Usually, when the PDM is fully charged, it is available for use of one week
(7 days).
⚫
Blue light flashes when PDM is charging, and the green light is always on
indicates full charge.
⚫
Only person (including patient) with adequate training is permitted to
operate the PDM.
Charging process:
1. Connect PDM to adapter.
2. Plug the adapter into a power socket.
B ureau Veritas
Содержание TouchCare A7+
Страница 1: ...B u r e a u V e r i t a s ...
Страница 24: ...13 B u r e a u V e r i t a s ...
Страница 25: ...14 B u r e a u V e r i t a s ...
Страница 29: ...18 B u r e a u V e r i t a s ...
Страница 81: ...70 B u r e a u V e r i t a s ...
Страница 125: ...114 B u r e a u V e r i t a s ...