5. To provide VVI Backup pacing during an atrial therapy, select VVI Backup… and set
the VVI Backup pacing parameters.
6. Select [DELIVER].
7. If necessary, select [ABORT] to terminate the manual therapy.
10.8.3 Operation of manual therapies
In general, each manual therapy with a corresponding automatic therapy performs in the
same manner as its automatic counterpart.
Antitachycardia pacing therapies – Manual ATP therapies deliver one sequence of the
selected therapy. For information about the operation of atrial Ramp and Burst+ therapies,
see Section 9.5, “Treating AT/AF episodes with antitachycardia pacing”, page 396. For
information about the operation of ventricular Ramp, Burst, and Ramp+ therapies, see
Section 9.2, “Treating VT and FVT episodes with antitachycardia pacing therapies”,
page 368.
Defibrillation – Manual defibrillation therapy charges the device capacitors and delivers a
biphasic shock, which is synchronized to a sensed R-wave if possible. The device does not
confirm the presence of VF before delivering the shock. For more information about
defibrillation, see Section 9.1, “Treating episodes detected as VF”, page 356.
Ventricular cardioversion – Manual ventricular cardioversion therapy charges the
capacitors and attempts to synchronize the shock to a ventricular sensed event that is
outside the refractory period. If the device cannot synchronize the therapy, the device aborts
the therapy. For more information, see Section 9.3, “Treating VT and FVT with ventricular
cardioversion”, page 380.
Atrial cardioversion – Manual atrial cardioversion therapy charges the capacitors and
attempts to synchronize the shock to a ventricular sensed event that is outside of the
refractory period. If the ventricular interval is shorter than the selected R-R Interval, the
device aborts the therapy. For more information, see Section 9.6, “Treating AT/AF with atrial