![Medtronic PROTECTA XT DR D314DRG Скачать руководство пользователя страница 444](http://html.mh-extra.com/html/medtronic/protecta-xt-dr-d314drg/protecta-xt-dr-d314drg_clinician-manual_1763148444.webp)
1. Select Tests > EP Study.
2. Select PES from the list of inductions and therapies.
3. If the Select Chamber dialog box appears, select [Atrium] or [RV].
4. If you want to treat the induced episode with a manual therapy, select [Suspend] to
prevent automatic detection.
Note: The [Suspend] button is located at the top of the screen and is not shown in
5. Select the Resume at DELIVER check box for automatic detection and therapy, or
clear the check box for manual therapy.
6. Accept the displayed test values or select new test values.
7. If you want to provide VVI Backup pacing during an atrial induction, select values for
VVI Backup.
8. Select [DELIVER PES].
9. If necessary, select [ABORT] to abort a therapy in progress.
10.8 Delivering a manual therapy
Manual therapies are tachyarrhythmia therapies you initiate from the programmer. During
EP testing, you can use manual therapies to provide backup therapy. At follow-up
appointments, manual therapies may be helpful in assessing therapy effectiveness and
making any necessary adjustments as part of chronic care.
The available manual therapies are Defibrillation, Cardioversion, Ramp, Burst, Ramp+, and
Burst+.
10.8.1 Considerations
Warning: Monitor the patient carefully when delivering a manual therapy. Have an external
defibrillator nearby and ready for immediate use. Potentially harmful tachyarrhythmias may
occur during device testing.
Aborting an induction or a therapy – As a safety precaution, you can select the
[ABORT] button displayed on the programmer to immediately terminate any induction,
manual or automatic therapy in progress. When a manual therapy is delivered, the device
automatically aborts any induction or automatic therapy already in progress.