10.7 Inducing an arrhythmia
The device provides several electrophysiology study (EP study) functions, including cardiac
stimulation protocols that induce tachyarrhythmias. The available induction methods are
T-Shock, 50 Hz Burst, Fixed Burst, and PES. These induction protocols may be used to
induce arrhythmias during EP testing to evaluate the effectiveness of tachyarrhythmia
therapies.
10.7.1 Considerations for inducing an arrhythmia
Warning: Monitor the patient carefully when using an EP study function. Have an external
defibrillator ready for use when inducing any tachyarrhythmia. An induced tachyarrhythmia
may degenerate to ventricular fibrillation.
Telemetry link – Make sure that there is a telemetry link between the device and the
programmer before performing an EP study function. Successful interrogation or
programming confirms proper communication between the device and the programmer.
When using wireless telemetry, verify that at least 3 of the green lights on the wireless
telemetry icon are illuminated. For more information see Section 4.1, “Establishing telemetry
between the device and the programmer”, page 45.
Resuming detection – Tachyarrhythmia detection is automatically suspended during all
EP study functions. Following a manual therapy, detection must be resumed manually.
Following an induction, detection is resumed either automatically or manually.
●
Following EP study manual therapies, detection remains suspended until [Resume] is
selected or until the programming head is removed from the implanted device.
●
Following EP study inductions, detection is resumed automatically, with one exception:
If [Suspend] was selected and the Resume at BURST or Resume at DELIVER check
box was NOT selected before performing the induction, detection remains suspended
until [Resume] is selected or until the programming head is removed from the implanted
device.
Note: The SUSPENDED annotation is only displayed on the programmer status bar during
EP study inductions if [Suspend] was selected before performing the induction.
Aborting an induction or therapy – As a safety measure, the programmer displays an
[ABORT] button that may be selected to immediately abort any induction or tachyarrhythmia
therapy in progress. A burst induction may also be aborted by removing the touch pen from
the [Press and Hold] button. When a manual therapy is delivered, the device automatically
aborts any induction or automatic therapy in progress.
Temporary parameter values – The EP study functions use test values that do not change
the programmed parameters of the device. The test values take effect when the induction
Medtronic
PROTECTA™ XT DR D314DRG
436
Clinician Manual