10.3.1 Evaluating the current template
You can use the Wavelet test to evaluate the accuracy of the current Wavelet template. As
the test is conducted, intrinsic QRS waveforms are assigned match scores (percentages)
in the Live Rhythm Monitor area. The higher the percentage, the closer the template event
matches the patient’s intrinsic event. Waveforms that fall below the programmable Match
Threshold value are determined to be non-match events.
Note that changes to the Match Threshold value may adversely affect Wavelet operation.
For information about Wavelet and the impact of increasing or decreasing the Match
Threshold value, refer to Section 8.4, “Discriminating VT/VF from SVT using Wavelet”,
10.3.1.1 Considerations for evaluating a template
Select temporary pacing settings – To increase the likelihood that sensed events will
occur during the test, you can select temporary pacing settings that evoke the patient’s
intrinsic rhythm.
Patient comfort – Reduce the pacing rate gradually to minimize symptoms associated with
abrupt changes in heart rate.
DOO, VOO and AOO pacing modes – The Wavelet test cannot be performed if the
programmed pacing mode is AOO, VOO, or DOO, because sensing is turned off in these
modes.
Tachyarrhythmia detection suspended – Tachyarrhythmia detection is suspended
when you select [Show Match Scores] to evaluate a Wavelet template.
10.3.1.2 How to evaluate the current template
1. Interrogate the device.
2. Select Tests > Wavelet.
3. Set the Mode, AV Delay, and Lower Rate for the test, or accept the values displayed.
4. Select [SHOW Match Scores].
Medtronic
PROTECTA™ XT DR D314DRG
428
Clinician Manual