Figure 213. Overview of patient-activated atrial cardioversion (CV)
Patient requests atrial CV.
An AT/AF episode is in
progress.
The delivery conditions are
met.
Therapy is pending. Charge
capacitors.
Synchronize and deliver
therapy or abort therapy.
Patient-activated atrial cardioversion uses the same methods for synchronization and
therapy delivery as automatic atrial cardioversion. For more information, refer to Section 9.6,
“Treating AT/AF with atrial cardioversion”, page 407.
9.7.2.1 Delivery conditions for patient-activated atrial cardioversion
For a patient-activated atrial cardioversion to become pending, the following conditions
must be met:
●
The device is programmed to allow patient-activated atrial cardioversion therapy.
●
The request for cardioversion therapy is received during an AT/AF episode.
●
A ventricular episode or VT Monitor episode is not in progress.
●
The atrial episode length is shorter than the programmed Duration to Stop therapy
parameter.
●
The atrial therapies have not been disabled.
Medtronic
PROTECTA™ XT DR D314DRG
Clinician Manual
415