2 The first 50 Hz Burst sequence is delivered for a programmed duration. This sequence fails to
terminate the AT/AF episode.
3 The device redetects the AT/AF episode.
4 An identical 50 Hz Burst sequence is delivered. This sequence terminates the AT/AF episode.
9.5.2.6 Ventricular backup pacing during an atrial ATP therapy
Ventricular backup pacing in the VVI and VOO modes is available during atrial ATP therapy
delivery. The backup pacing is delivered either at the separately programmed Lower Rate
or at the current pacing rate, whichever is faster. The backup pacing output is preset at 6 V
and 1.5 ms.
The following options are available for enabling VVI/VOO Backup Pacing:
●
On (Always): backup pacing is delivered during every atrial ATP therapy.
●
On (Auto Enable): backup pacing is delivered if 1 of the 4 ventricular events preceding
the therapy is paced. When Auto Enable is selected, the device monitors for rapidly
conducting ventricular sense events that may occur during an ATP therapy delivery.
Note: VVI Backup Pacing could be competitive with intrinsic ventricular activity during the
atrial ATP sequence.
9.5.2.7 Automatically disabling atrial therapies
In some situations the device may automatically disable or suspend an ATP therapy.
VT/VF detection after an AT/AF therapy delivery – Atrial therapies are disabled if VT/VF
is detected immediately after an AT/AF therapy is delivered. In this case, atrial therapies
remain disabled until you reprogram them.
VT/VF detection unrelated to AT/AF therapy delivery – If the device detects VT/VF
during an AT/AF episode, but the detection is not related to therapy delivery, it temporarily
suspends atrial therapies. Atrial therapies automatically resume when the VT/VF episode
ends.
For information about programmable options that disable therapies, refer to Section 9.4,
“Scheduling atrial therapies”, page 389.
9.5.3 Programming considerations for atrial ATP therapies
Backup pacing for 50 Hz Burst therapy – VOO backup pacing is competitive with the
intrinsic ventricular rate if there is an intrinsic rate.
VF therapy – You must program VF therapy to On before programming atrial ATP therapies
to On.