![Medtronic PROTECTA XT DR D314DRG Скачать руководство пользователя страница 377](http://html.mh-extra.com/html/medtronic/protecta-xt-dr-d314drg/protecta-xt-dr-d314drg_clinician-manual_1763148377.webp)
9.2.4.3 Programming Ramp+ pacing therapy
Select Params icon
⇒ VT | Therapies…
▷ VT Therapy Status <On>
▷ Therapy Type <Ramp+>
▷ Initial # Pulses
▷ R-S1 Interval=(%RR)
▷ S1S2 (Ramp+)=(%RR)
▷ S2SN (Ramp+)=(%RR)
▷ # Sequences
▷ Smart Mode
Note: Similar programming steps apply for Ramp+ therapies for FVT episodes.
9.2.4.4 Programming Shared Settings
Select Params icon
⇒ VT | Therapies…
⇒ Shared Settings…
▷ V-V Minimum ATP Interval
▷ V. Amplitude
▷ V. Pulse Width
▷ V. Pace Blanking
9.2.5 Evaluation of ventricular ATP therapies
9.2.5.1 The Quick Look II screen
The Quick Look II screen provides information about VT/VF therapies.
Select Data icon
⇒ Quick Look II
Treated VT/VF episodes – This section includes a count of treated VT/VF episodes. You
can select the Treated [>>] button to view the data for treated episodes.
Quick Look II observations – The Quick Look II observations are based on an analysis
of interrogated data since the last session and programmed parameters. If related
Medtronic
PROTECTA™ XT DR D314DRG
Clinician Manual