Figure 183. Operation of ChargeSaver and Switchback
ATP During Charging
ATP Before Charging
ChargeSaver
Switchback
Smart Mode feature – If the Smart Mode feature is programmed to On, the device
automatically sets the ATP parameter to Off if ATP therapies delivered before or during
charging fail to terminate the tachyarrhythmia in 4 consecutive episodes.
9.1.2.4 Confirming VF for the first defibrillation
Before the device delivers the first defibrillation shock for a VF episode, it monitors cardiac
rhythm to confirm the presence of VF. VF is confirmed using one of the following
confirmation intervals:
●
an interval calculated from the ventricular cycle 60 ms, if this interval is at least
as long as the programmed VF detection interval. The interval is provided by the
Confi option (nominally programmed to On) and is used when the ventricular
rhythm is stable.
●
the programmed VT In 60 ms (or the programmed VF Interval if VT Detection is
programmed to Off). This interval is used if Confi is turned off.
The device classifies any ventricular event that occurs within the confirmation interval as an
“arrhythmic event” and any event that occurs outside the interval as a “normal event”.
With each ventricular event, the device reviews the previous 5 ventricular events. If the
previous 5 ventricular events include 4 “normal events”, the device aborts the therapy.
Medtronic
PROTECTA™ XT DR D314DRG
360
Clinician Manual