Zone merging – To ensure that the device delivers sufficiently aggressive therapies when
FVT Detection is programmed, the device merges detection zones during redetection, as
follows:
●
If FVT Detection is programmed to FVT via VT and an FVT or VF episode is detected,
the VT zone merges with the FVT zone. After the zones have merged, the episode
cannot be classified for redetection as the slower VT rhythm.
●
If FVT Detection is programmed to FVT via VF and a VF episode is detected, the FVT
zone merges with the VF zone. After the zones have merged, the episode cannot be
classified for redetection as the slower FVT rhythm.
The merged zone configuration remains in effect until the episode terminates.
8.2.2.9 Evaluating the ventricular rhythm for termination
The device determines that a VT episode has terminated if 8 consecutive ventricular
intervals are longer than or equal to the programmed VT detection interval, or if 20 s elapse
during which the median of the last 12 ventricular intervals is always longer than the VT
detection interval.
If VT detection is off but VF detection is on, the device uses the programmed VF interval to
terminate a VF episode, and the episode is terminated when 8 consecutive ventricular
intervals are longer than or equal to the programmed VF detection interval, or when 20 s
elapse during which the median of the last 12 ventricular intervals is always longer than the
VF detection interval.
8.2.2.10 SVT discrimination features for ventricular tachyarrhythmia
detection
The device provides the following features designed to help prevent conducted
supraventricular tachycardias (SVTs) from being treated as ventricular tachyarrhythmias:
PR Logic – PR Logic is a set of SVT discrimination functions that withhold VT/VF detection
if the rhythm displays characteristics of an SVT origin. For more information, see Section 8.3,
“Discriminating VT/VF from SVT using PR Logic”, page 321.
Wavelet – The Wavelet feature withholds VT/VF detection if the rhythm displays
characteristics of an SVT origin. For more information, see Section 8.4, “Discriminating
VT/VF from SVT using Wavelet”, page 326.
Onset – The Onset feature is designed to prevent sinus tachycardia from being treated as
ventricular tachycardia. For more information, see Section 8.5, “Discriminating sinus
tachycardia from VT using the Onset feature”, page 334.
Medtronic
PROTECTA™ XT DR D314DRG
Clinician Manual
313