Figure 146. Initial Beats to Detect calculation for VT
A
S
A
S
V
S
V
S
V
S
V
S
A
R
A
R
A
R
A
R
A
R
A
R
A
R
1
0
1
2
3
4
5
T
S
T
S
T
S
T
S
T
S
T
S
V
S
A
R
A
R
A
R
A
R
A
R
A
R
A
R
A
R
A
R
A
R
A
R
A
R
A
R
7
6
8
9
10
11
12
13
14
15
16
200 ms
T
S
T
S
T
S
T
S
T
S
T
S
T
S
T
S
T
S
T
S
T
D
VT event count
ECG
VT interval
VT event count
ECG
VT interval
Marker Channel
Marker Channel
1 Ventricular tachycardia starts, and sensed ventricular intervals in the VT detection zone are
classified as VT events (marked TS).
2 A sensed ventricular interval occurs outside the VT detection zone. VT detection restarts.
3 The programmed VT Initial Beats to Detect value of 16 events is reached, and the device detects
VT (indicated by the TD marker).
8.2.2.3 Detecting 2 clinical VTs
A fast ventricular tachyarrhythmia (FVT) detection zone may be used to allow different
therapeutic approaches for a patient who exhibits 2 VTs of different rates. To detect 2 clinical
VTs, program FVT Detection to via VT and select a V. Interval (Rate) value for FVT. To help
ensure that the patient’s fast ventricular tachycardia is classified as FVT, select a value that
matches the longest ventricular interval that typically occurs during the patient’s fast VT.
Medtronic
PROTECTA™ XT DR D314DRG
308
Clinician Manual