8.1.5.3 Flashback Memory
Select Data icon
⇒ Clinical Diagnostics
⇒ Flashback Memory
The Flashback Memory screen shows interval and marker data prior to the most recent
occurrence of an AT/AF episode. Total elapsed time is plotted against interval length in ms.
Figure 138. Flashback Memory screen
8.1.5.4 Cardiac Compass Report
Select Data icon
⇒ Clinical Diagnostics
⇒ Cardiac Compass Trends (Report Only)
The Cardiac Compass Report provides information about AT/AF episodes and ventricular
rhythms, and how much time the patient has spent in AT/AF.
The V. rate during AT/AF trend on the Cardiac Compass Report displays information about
ventricular response during atrial tachyarrhythmias.