8.1.2.6 Monitoring an atrial tachyarrhythmia without delivering therapy
When atrial tachyarrhythmia detection is programmed to Monitor, the device does not
deliver AT/AF therapies, and there is no redetection. All other operations, including Mode
Switch, remain unchanged.
8.1.3 Programming considerations for AT/AF detection
VF detection backup during AT/AF – For AT/AF Detection to be programmed to On, VF
Detection must be programmed to On. This ensures VF detection backup during AT/AF
episodes.
Asynchronous pacing mode – AT/AF Detection cannot be programmed to On when the
programmed pacing mode is DOO, VOO, or AOO.
8.1.4 Programming AT/AF detection
8.1.4.1 Programming AT/AF detection
Select Params icon
▷ AT/AF Detection
▷ AT/AF Interval (Rate)
8.1.4.2 Programming AT/AF detection for 2 detection zones
Select Params icon
⇒ AT/AF | Therapies…
▷ Detection <On>
▷ Zones <2>
▷ Fast AT/AF A. Interval (Rate)
▷ AT/AF A. Interval (Rate)
8.1.5 Evaluation of AT/AF detection
8.1.5.1 Quick Look II screen
Select Data icon
⇒ Quick Look II
The Quick Look II screen shows the total percentage of time that the patient has spent in
AT/AF and the number of monitored or treated AT/AF episodes since the last session.