Ablation (RF ablation or microwave ablation) – Ablation is a surgical technique in which
radio frequency (RF) or microwave energy is used to destroy cells by creating heat. Ablation
used in cardiac device patients may result in, but is not limited to, induced ventricular
tachyarrhythmias, oversensing, unintended tissue damage, device damage, or device
malfunction.
Pulse-modulated ablation systems may pose higher risk for induced ventricular
tachyarrhythmias. Medtronic cardiac devices are designed to withstand exposure to
ablation energy. To mitigate risks, observe the following precautions:
●
Ensure that temporary pacing and defibrillation equipment is available.
●
Avoid direct contact between the ablation catheter and the implanted system.
●
Position the return electrode patch so that the electrical current pathway does not pass
through or near the device and leads.
●
Always monitor the patient during ablation with at least two separate methods, such as
arterial pressure display, ECG, manual monitoring of the patient’s rhythm (taking pulse)
or monitor by some other means such as ear or finger pulse oximetry, or Doppler pulse
detection.
To avoid or mitigate the effects of oversensing, if appropriate for the patient, initiate
asynchronous pacing by implementing one of the following precautions;
●
Suspend tachyarrhythmia detection by using a magnet or a programmer. If a
programmer is used and ablation causes a device reset, the cardiac device resumes
detection. After the ablation procedure, remove the magnet or restore device
parameters.
●
If appropriate for the patient, program the device to an asynchronous pacing mode (for
example, DOO). After the ablation procedure, remove the magnet or restore device
parameters.
Capsule endoscopy, pH capsule procedures – Capsule endoscopy is a procedure in
which a capsule containing a tiny camera is swallowed by the patient to take pictures of the
patient’s digestive tract. Capsule endoscopy and pH capsule procedures should pose no
risk of electromagnetic interference.
Dental procedures – Dental equipment, such as ultrasonic scalers, drills, and pulp testers,
poses no risk of electromagnetic interference. Keep a cardiac device at least 15 cm (6 in)
away from magnets, such as magnets found in dental office pillow headrests.
Diagnostic radiology (CT scans, fluoroscopy, mammograms, x-rays) – Diagnostic
radiology refers to the following medical procedures:
●
Computed axial tomography (CT or CAT scan)
●
Fluoroscopy (an x-ray procedure that makes it possible to see internal organs in motion
by producing a video image)
Medtronic
PROTECTA™ XT DR D314DRG
Clinician Manual
29