7.17.10 Programming PMOP
Select Params icon
⇒ Pacing…
⇒ Arrhythmia/Post Shock…
▷ Post Mode Switch <On>
▷ Overdrive Rate
▷ Overdrive Duration
7.17.11 Evaluation of atrial intervention pacing
The device collects and stores AT/AF episode summary data that includes the total
percentage of time that the device provided atrial intervention pacing. You can view the
AT/AF summary data on the programmer screen and print the data in report form. For more
information, see Section 6.9, “Viewing episode and therapy counters”, page 175.
Select Data icon
⇒ Clinical Diagnostics
⇒ Counters
⇒ AT/AF Episodes
The % of Time Atrial Intervention line in the AT/AF Summary section of the Data-Counters
screen displays the total percentage of time that the patient received atrial intervention
pacing. The displayed percentage reflects the combined total of pacing resulting from ARS
and APP.
Figure 124. Example of the AT/AF Summary section of the Data-Counters screen
Note: If APP is enabled, atrial intervention pacing is more likely to have resulted from APP
than ARS or PMOP.