Figure 120. Atrial Preference Pacing (APP)
A
tri
al
ra
te
min
-1
(bp
m
)
Time (s)
Scheduled pace
Atrial Preference Pacing
Search Beats = 10
Rate increase
55
0
5
10
15
20
25
30
60
65
70
75
Intrinsic beats
Atrial Preference Pacing
Atrial rate
APP is a programmable feature that is designed to maximize atrial overdrive pacing when
the patient is not experiencing an atrial tachyarrhythmia. The device responds to changes
in the atrial rate by accelerating the pacing rate until reaching a steady paced rhythm that
is slightly faster than the intrinsic rate.
After each nonrefractory atrial sensed event, the device decreases the atrial pacing interval
by the programmed Interval Decrement value. This progression continues until the pacing
rate exceeds the intrinsic rate, resulting in an atrial paced rhythm. It sustains this increased
rate for the number of beats programmed for a Search Beats parameter, then decreases
the pacing rate slightly (by 20 ms) to search for the next intrinsic beat. This results in a
dynamic, controlled, stair-step increase or decrease in the pacing interval resulting in a
pacing rate slightly above the intrinsic rate. The Maximum Rate parameter sets an upper
rate limit for APP.
Atrial pacing pulses delivered for APP are annotated on the Marker Channel with PP
(proactive pace).