Figure 108. Operation of Rate Hysteresis in VVI mode
V
P
V
P
V
S
V
S
V
P
V
P
200 ms
A
S
A
S
A
S
A
S
A
S
Lower Rate
interval
ECG
Marker Channel
1 The device paces in VVI mode at the programmed Lower Rate.
2 After a ventricular sensed event, the device applies the hysteresis interval (shaded bar).
3 A sensed event occurs before the hysteresis interval expires, so hysteresis operation continues.
4 The hysteresis interval expires, and the device paces the ventricle and reapplies the Lower Rate
interval.
5 The ventricle is paced at the Lower Rate.
7.9.3 Programming considerations for Rate Hysteresis
Verifying adequate cardiac support – The programmed hysteresis rate determines the
slowest heart rate that can occur before pacing starts. Ensure that the selected hysteresis
rate is adequate to support the patient’s cardiac condition.
Programming the hysteresis rate – To avoid large, sudden changes in heart rate, you
would normally select a hysteresis rate that is no more than 30 bpm below the programmed
Lower Rate.
Lower Rate – You cannot program the hysteresis rate to a value equal to or above the
Lower Rate.
Compatibility – Rate Hysteresis cannot be enabled at the same time as Ventricular Rate
Stabilization, Atrial Rate Stabilization, or Atrial Preference Pacing.