Figure 103. Drop Detection
Detection window: 1
minute
Drop Size: 25 min
-1
(bpm)
Intervention Rate: 100 min
-1
(bpm)
Drop Rate: 60 min
-1
(bpm)
Lower Rate: 45 min
-1
(bpm)
Sensed beats
Paced beats
With Drop Detection, the device intervenes when the ventricular rate drops by a specified
number of beats per minute to below a specified heart rate within a specified period of time.
These conditions are established by programming the Drop Size, Drop Rate, and Detection
Window parameters, respectively.
Figure 104. Low Rate Detection
Intervention Rate:
100 min
-1
(bpm)
Lower Rate: 45 min
-1
(bpm)
Paced beats
Sensed beats
Detection beats
With Low Rate Detection, the device intervenes when the atrium is paced at the Lower Rate
for the number of consecutive beats specified by the Detection Beats parameter.
Medtronic
PROTECTA™ XT DR D314DRG
258
Clinician Manual