7.7.4 Programming Auto PVARP
Select Params icon
⇒ Pacing…
⇒ PVARP…
▷ PVARP <Auto>
▷ Minimum PVARP
7.8 Treating syncope with Rate Drop Response
Patients with carotid sinus syndrome or vasovagal syncope may lose consciousness or
experience related symptoms after significant heart rate drops. When syncope is caused
primarily by cardioinhibition and when permanent AF is not present, pacing at an elevated
rate may prevent syncope and related symptoms from occurring.
7.8.1 System solution: Rate Drop Response
Rate Drop Response monitors the heart for significant rate drops and responds by pacing
the heart at an elevated rate.
7.8.2 Operation of Rate Drop Response
Figure 101. Overview of Rate Drop Response
Sinus rate
Intervention
Termination
Step - down
Detection
Rate Drop Response operates in phases. During the detection phase, the device monitors
the heart for rate drops that conform to programmed criteria. During the intervention phase,
the device paces the heart at a programmed elevated rate for a programmed duration.
During the step-down phase, the device gradually slows pacing to the sinus rate or the
Lower Rate.
Medtronic
PROTECTA™ XT DR D314DRG
256
Clinician Manual