Figure 90. Overview of Capture Management
Pacing
threshold
search
Pacing output
adjustment
Pacing output
determination
Collect
threshold trend
data
(if programmed)
7.5.2.1 Manual adjustment of pacing outputs
You have the option to program pacing outputs manually instead of using automatic Capture
Management. The pacing safety margins should be checked if Capture Management is
programmed to the Monitor setting. Threshold data that is collected during pacing threshold
searches can make it easier for you to select values for pacing output parameters. For more
information about manual programming, refer to Section 7.2, “Providing pacing therapies”,
7.5.2.2 Pacing thresholds and safety margins
The amplitude and pulse width parameters control the output energy of pacing pulses in
each chamber. The pacing output energy determines whether pacing pulses capture the
myocardium. It is necessary for pacing output settings to exceed the pacing threshold by a
safety margin. Pacing threshold variations may be caused by exercise, eating, sleeping,
drug therapy, or changes in other cardiac conditions.
Both a threshold curve and a safety margin curve are shown in Figure 91. The threshold
curve consists of combinations of amplitude and pulse width settings. Pacing output settings
on or above the curve result in capture, whereas settings below the curve result in loss of
capture. The safety margin curve consists of pacing output settings, each of which has a
target amplitude that is equal to a threshold amplitude with a safety margin applied.
Medtronic
PROTECTA™ XT DR D314DRG
Clinician Manual
239