Manual Rate Response (Rate Profile Optimization Programmed to Off) – With manual
Rate Response, the rate curve is established during a patient session when the rates and
setpoints are programmed. The rate curve remains constant until the parameters are
reprogrammed.
7.4.2.3 Rate Profile Optimization
Rate Profile Optimization automatically adjusts the patient’s rate response between office
visits. The goal of Rate Profile Optimization is to ensure that the rate response remains
appropriate for the full range of patient activities. Each day, the device collects and stores
daily and long-term averages of the percentage of time that the patient sensor indicated
rate is at different pacing rates. The device then uses the ADL Response and Exertion
Response parameters to define the percentage of time that the pacing rate stays in the ADL
rate range and exertion rate range respectively. Based on daily comparisons, the device
adjusts either the ADL Setpoint, the UR Setpoint, or both setpoints.
By programming new settings for rates or Rate Profile Optimization, you are affecting the
comparisons. Immediate changes occur. These changes project how rate response should
change in the future based on stored sensor rate information and the selected Rate Profile
Optimization settings. The device continues to adjust the rate response over time.
The device adapts Rate Response more rapidly for the first 10 days after Rate Profile
Optimization is first activated post-implant or after certain Rate Response parameters are
manually reprogrammed (Lower Rate, ADL Rate, Upper Sensor Rate, ADL Response, or
Exertion Response). The intent is to quickly match Rate Response to the operation
prescribed by the parameter changes.
Note: Because the device is automatically changing the setpoint values, if you manually
program the setpoint values, Rate Profile Optimization is disabled.
7.4.2.4 Activity Acceleration and Activity Deceleration
The Activity Acceleration and Activity Deceleration functions are used to smooth the pacing
rate. Activity Acceleration controls how rapidly the pacing rate increases. Activity
Deceleration controls how rapidly the pacing rate decreases and has both fixed values and
the “Exercise” option. The Exercise setting adjusts the deceleration dynamically based on
the intensity and duration of exercise, and it can extend the deceleration up to 20 min.
As shown in Figure 86, changing the values of the Activity Acceleration and Activity
Deceleration parameters affects the pacing rate during and after exertion.
Medtronic
PROTECTA™ XT DR D314DRG
Clinician Manual
233