7.1.8.2 Viewing the Sensing Integrity Counter
Select Data icon
⇒ Device/Lead Diagnostics
⇒ Battery and Lead Measurements
Figure 68. Battery and Lead Measurements screen
The Sensing Integrity Counter records the following information:
●
the number of short ventricular intervals that occurred since the last patient session
●
the date of the first of these intervals, if any occurred
A large number of short ventricular intervals may indicate oversensing, lead fracture, or a
loose setscrew.
Note: The number of short ventricular intervals is an input to the RV Lead Integrity Alert.
For more information, see Section 6.4, “Monitoring leads using RV Lead Integrity Alert”,
Medtronic
PROTECTA™ XT DR D314DRG
Clinician Manual
211