7.1.7 Programming sensing
7.1.7.1 Programming sensitivities, polarities, and blanking periods
Select Params icon
⇒ Pacing…
▷ A. Sensitivity
▷ RV Sensitivity
▷ RV Sense Polarity
⇒ Blanking…
▷ PVAB Interval
▷ PVAB Method
▷ A. Blank Post AP
▷ A. Blank Post AS
▷ V. Blank Post VP
▷ V. Blank Post VS
7.1.7.2 Programming refractory periods
Select Params icon
⇒ Pacing…
⇒ PVARP…
▷ PVARP (or A. Refractory)
▷ Minimum PVARP
7.1.8 Evaluation of sensing
7.1.8.1 Using the Sensing Test to evaluate sensing
The Sensing Test allows you to measure P-wave and R-wave amplitudes. These
measurements may be useful for assessing lead integrity and sensing performance. After
the Sensing Test is complete, the test results are displayed on the test screen. You may
view and print the results when desired. For more information, refer to Section 10.5,
“Performing a Sensing Test”, page 432.
Medtronic
PROTECTA™ XT DR D314DRG
210
Clinician Manual