The availability of refractory periods depends on the programmed pacing mode. The Post
Ventricular Atrial Refractory Period (PVARP) is available in dual chamber pacing modes,
and the Atrial Refractory Period is available in atrial pacing modes.
7.1.5.1 Post Ventricular Atrial Refractory Period (PVARP)
The Post Ventricular Atrial Refractory Period (PVARP) follows a paced, sensed, or refractory
sensed ventricular event. An atrial event that is sensed during this interval is classified as
a refractory event. It does not inhibit a scheduled atrial pace or start a Sensed AV interval.
The PVARP setting is only programmable for dual chamber pacing modes (except DOO
mode).
●
When the device is operating in the DDDR and DDD modes, the PVARP setting
prevents the tracking of retrograde P-waves that could initiate a pacemaker-mediated
tachycardia.
●
When the device is operating in the DDIR and DDI modes, the PVARP setting prevents
the inhibition of atrial pacing based on sensed retrograde P-waves. PVARP should be
programmed to a value longer than the VA interval (retrograde) conduction time.
Figure 67. Timing for fixed PVARP
A
P
A
P
V
P
V
P
Marker Channel
PVAB
Fixed
PVARP
The PVARP parameter may be programmed to Auto instead of a fixed value. Auto PVARP
adjusts PVARP in response to changes in the patient’s intrinsic rate or pacing rate. During
a Mode Switch episode, the device enables Auto PVARP. For more information, see
Section 7.7, “Adjusting PVARP to changes in the patient’s heart rate”, page 253.
The PVARP setting may be extended by the PVC Response feature or the PMT Intervention
feature.
7.1.5.2 Atrial Refractory Period
The Atrial Refractory Period setting is programmable only for the AAI and AAIR single
chamber pacing modes. The Atrial Refractory Period prevents the inhibition of atrial pacing
due to sensed far-field R-waves or noise.
Medtronic
PROTECTA™ XT DR D314DRG
Clinician Manual
207