6.15.5 Considering how diagnostic features with data storage impact
longevity
Pre-arrhythmia EGM storage – Continual use of Pre-arrhythmia EGM storage reduces
device longevity. For a patient with uniform tachyarrhythmia onset mechanisms, the
greatest benefit of Pre-arrhythmia EGM storage is obtained after capturing a few episodes.
When Pre-arrhythmia EGM storage is on, the device collects up to 10 s of EGM data before
the onset of VT/VF, VT Monitor, or the detection of SVT episodes. The device stores up to
5 s of EGM before AT/AF detection regardless of the Pre-arrhythmia EGM storage setting.
To balance the benefit of using the Pre-arrhythmia EGM storage feature with optimizing
device longevity, consider the following programming options:
●
Set Pre-arrhythmia EGM storage to On to capture possible changes in the
tachyarrhythmia onset mechanism following significant clinical adjustments such as
device implant, medication changes, and surgical procedures. Pre-arrhythmia EGM
storage may be set to On-1 month, On-3 months, or On-continuous. Select the setting
for the shortest time period that will provide the necessary data.
●
Set Pre-arrhythmia EGM storage to Off after you have obtained the data of interest.
Note: When Pre-arrhythmia EGM storage is set to Off, the device begins to store EGM
information for VT/VF, VT Monitor, and SVT episodes after the third tachyarrhythmia event
occurs. Though EGM is not recorded before the start of the arrhythmia, the device still
records up to 20 s of data before the onset or detection of the episode. This data includes
interval measurements and Marker Channel annotations. In addition, Flashback Memory
data is stored for the most recent tachyarrhythmia episodes.
Holter Telemetry – Extended use of the Holter Telemetry feature decreases battery
longevity. The Holter Telemetry feature continues to transmit EGM and Marker channel data
for the programmed time duration regardless of whether the programming head is
positioned over the device.
Medtronic
PROTECTA™ XT DR D314DRG
200
Clinician Manual