ATP Sequences – The number of atrial ATP sequences that were delivered and the number
that were aborted.
Automatic Shocks – The number of automatic atrial defibrillation therapies that were
delivered and the number that failed to terminate.
Patient Activated Shocks – The number of patient-activated atrial defibrillation therapies
that were delivered and the number that failed to terminate.
6.10 Viewing Flashback Memory data
Flashback Memory records atrial and ventricular intervals that occur immediately prior to
tachyarrhythmia episodes or the most recent interrogation. The feature plots the interval
data over time and allows you to view and print a graph of the collected data. The graphed
data may help you assess the patient’s heart rhythm and the performance of other features
such as Rate Response.
Flashback Memory automatically records up to a total of 2000 V-V and A-A intervals and
stored marker data for the following events:
●
the most recent interrogation
●
the most recent VF episode
●
the most recent VT episode
●
the most recent AT/AF episode
If 2 or more episodes are detected within 15 min, the Flashback Memory data before the
episodes may be truncated.
Note: When an episode is detected, Flashback Memory storage is suspended until the
episode terminates.
6.10.1 How to view Flashback Memory data
Select Data icon
⇒ Clinical Diagnostics
⇒ Flashback Memory
Note: You can also display the Flashback Memory screen by selecting [Flashback] from
the most recent VT, VF, FVT, or AT/AF record details screens.
Medtronic
PROTECTA™ XT DR D314DRG
180
Clinician Manual