6.9.2 VT/VF episode counters
Figure 51. VT/VF episode counters
The following count data is available for VT/VF episodes:
VF, FVT, and VT – The number of episodes of each tachyarrhythmia.
Monitored VT – The number of VT Monitor episodes.
VT-NS – The number of non-sustained ventricular tachyarrhythmias.
High Rate-NS – The number of high rate non-sustained ventricular tachyarrhythmia
episodes.
PVC Runs – The average number of runs per hour of premature ventricular contractions
(PVCs) in which 2, 3, or 4 consecutive ventricular events are premature.
PVC Singles – The average number of single PVCs per hour. PVCs in PVC runs are not
counted as PVC singles.
Runs of VRS Paces – The average number of times per hour that 2 or more consecutive
ventricular events are Ventricular Rate Stabilization (VRS) pacing pulses (VRS escape
interval timeouts).
Single VRS Paces – The average number of single VRS pacing pulses (VRS escape
interval timeouts) per hour. VRS paces in runs of VRS paces are not counted as single VRS
paces.