LECG source – You can choose to display the Leadless ECG waveform on the LECG
channel. Leadless ECG is an approximation of a surface ECG signal through the Can to
SVC source, and is available only when an SVC coil is present. You can also choose to
display the waveform from the RVcoil to Aring source on the LECG channel.
Note: The Leadless ECG option cannot be removed from the Live Rhythm Monitor.
For more information, see Section 4.12, “Expediting follow-up sessions with Leadless
ECG”, page 92.
EGM source – For each EGM channel, define the source electrodes between which the
device records EGM signals.
Note: The cardiac interval measurements of the device are always based on the signals
sensed through the programmed sensing polarity (not the stored diagnostic EGM).
Therefore, tachyarrhythmia interval criteria, synchronization, and therapy are not affected
by your selection of EGM sources.
EGM and LECG range – Select a range for each EGM channel and the LECG channel.
The range setting affects the resolution of the signal; the lower the setting, the higher the
resolution. If the signal is illegible or clipped, consider changing the range selection.
Monitored – Select a set of 2 sources to be used for episode record storage.
Pre-arrhythmia EGM – Indicate whether you want to store EGM data collected prior to an
episode. When Pre-arrhythmia EGM storage is on, the device collects up to 10 s of EGM
data before the onset of VT/VF, VT Monitor, or the detection of SVT episodes. If
Pre-arrhythmia EGM is programmed to Off, the episode record will store only intervals and
no EGM at the beginning of each VT/VF, VT Monitor, or SVT episode. The device stores
up to 5 s of EGM before AT/AF detection regardless of the Pre-arrhythmia EGM storage
setting.
Note: Pre-arrhythmia EGM storage works by keeping the EGM circuitry enabled at all times,
and therefore it reduces device longevity. If you select On - 1 Month or On - 3 Months,
Pre-arrhythmia EGM storage is automatically turned off after the time period expires.
Clearing data – The Clear data function clears all stored data except trend data and lifetime
counters.
Note: Cleared data is not recoverable.