Notes:
●
You should adjust the Reference Impedance only when all of the following conditions
hold: the patient has stable pulmonary fluid status, OptiVol trends show that the patient’s
Daily Impedance is stable, and the Reference Impedance has not already adjusted to
the patient’s Daily Impedance.
●
The Reference Impedance adjustment cannot be performed during the Reference
Impedance Initialization period.
●
The OptiVol observation is suspended for the first few days after an adjustment.
6.7.4 Programming OptiVol 2.0 Fluid Status Monitoring
6.7.4.1 Programming the OptiVol Threshold
Select Params icon
⇒ Alert…
⇒ OptiVol 2.0 Fluid Settings… | Device Tone
▷ OptiVol Threshold
6.7.4.2 Adjusting the Reference Impedance
Select Params icon
⇒ Alert…
⇒ OptiVol 2.0 Fluid Settings… | Device Tone
⇒ Additional Settings…
6.7.5 Evaluation of OptiVol 2.0 Fluid Status Monitoring
Caution: Verify lead integrity when evaluating OptiVol 2.0 Fluid Status Monitoring. Loss of
RVcoil integrity due to lead fracture or insulation defect may adversely affect the results of
OptiVol 2.0 Fluid Status Monitoring.
6.7.5.1 Viewing OptiVol 2.0 Fluid Trends
Select Reports icon
⇒ Heart Failure…
The Heart Failure Management Report and the Cardiac Compass Report display up to 14
months of OptiVol 2.0 Fluid Trends patient data.
Medtronic
PROTECTA™ XT DR D314DRG
164
Clinician Manual