OptiVol 2.0 Fluid Index – If the Daily Impedance falls below the Reference Impedance,
this may indicate that fluid is accumulating in the patient’s thoracic cavity. If the Daily
Impedance remains below the Reference Impedance, the difference between the Daily
Impedance and Reference Impedance values, adjusted for individual patient variation, is
added to the OptiVol 2.0 Fluid Index.
While there is a difference between the Daily Impedance and the Reference Impedance,
the fluid index may continue to increase. If the Daily Impedance begins to rise, this may be
an indication that the thoracic fluid accumulation is resolving and the fluid index may
decrease. When the Daily Impedance returns to the Reference Impedance, the fluid event
is considered to have ended and the OptiVol 2.0 Fluid Index resets to 0.
OptiVol Threshold – If the Daily Impedance remains below the Reference Impedance on
consecutive days, the OptiVol 2.0 Fluid Index may rise above the programmed OptiVol
Threshold. This triggers an OptiVol clinical status observation.
6.7.3 Programming considerations for OptiVol 2.0 Fluid Status Monitoring
Setting the OptiVol Threshold – The OptiVol Threshold is nominally programmed to 60.
Medtronic recommends that you use this setting until you have clinical experience using
OptiVol 2.0 Fluid Status Monitoring with individual patients.
If the patient is experiencing too many OptiVol observations, the OptiVol Threshold may be
set at too sensitive (low) a level, and you should consider increasing the OptiVol Threshold.
If OptiVol observations are absent or are delayed when the patient has thoracic fluid
accumulation, the OptiVol Threshold may be set at too insensitive (high) a level, and you
should consider decreasing the OptiVol Threshold.
Reference Impedance initialization period – The Reference Impedance is first
calculated on the thirty-fourth day of impedance measurements after implant. If the patient’s
lead is still maturing, the patient is retaining lung fluid, or there is tissue swelling around the
device pocket, the Reference Impedance may require more time to adapt to the patient’s
normal Daily Impedance.
Adjusting the Reference Impedance – Under appropriate circumstances, you can adjust
the Reference Impedance so that it more closely matches the patient’s Daily Impedance
measurements. This should be done only in rare cases and when the patient has stable
pulmonary fluid status. The adjustment process takes several days. The Reference
Impedance is set to the average of the last Daily Impedance measurement and the next 3
Daily Impedance measurements.