Heart rate variability – Reduced variability in the patient’s heart rate may help you to
identify heart failure decompensation. The device measures each atrial interval and
calculates the median atrial interval every 5 min. It then calculates and plots a variability
value (in ms) for each day.
Note: The heart rate variability calculation does not include events that occur during
arrhythmia episodes.
OptiVol 2.0 Fluid Index – A decrease in intrathoracic impedance may be an early indicator
of fluid accumulation associated with heart failure. The OptiVol 2.0 Fluid Index trend displays
the accumulated difference between the measured daily thoracic impedance and the
Reference Impedance, adjusted for individual patient variation. If the Daily Impedance is
less than the Reference Impedance, this may indicate that the patient’s thoracic fluid has
increased. The horizontal line shows the programmed value of the OptiVol threshold.
Caution: Verify lead integrity when evaluating the OptiVol 2.0 Fluid Index trend. Loss of
RVcoil integrity due to lead fracture or insulation defect may adversely affect the results of
the OptiVol 2.0 Fluid Index trend.
Note: The OptiVol Fluid Status Monitoring feature has been updated to OptiVol 2.0 to
account for individual patient variation, including allowing the Fluid Index to increase or
decrease based on recent intrathoracic impedance measurements.
For more information, see Section 6.7, “Monitoring for thoracic fluid accumulation with
Note: The OptiVol fluid monitoring feature provides an additional source of information for
patient management and does not replace assessments that are part of standard clinical
practice.
Thoracic impedance – The Thoracic impedance trend allows you to compare the daily
average measured thoracic impedance to the Reference Impedance values. The Reference
Impedance changes slightly from day to day, to slowly adapt to the Daily Impedance.
6.6 Viewing heart failure management information
An analysis of clinical information related to heart failure can help you to follow changes in
a patient’s condition and correlate these changes with variations in device programming,
medication, patient activity, or symptoms.