2. Check the Initial Interrogation Report or the Quick Look II screen to confirm that the
battery voltage is at least 3.0 V at room temperature.
If the device has recently delivered a high-voltage therapy or if the device has been
exposed to low temperatures, then the battery voltage will be temporarily lower and
the capacitor charge time may increase. Allow the device to warm to room temperature
and check the battery voltage again. If an acceptable battery voltage cannot be
obtained, contact a Medtronic representative.
3. Select Params > Data Collection Setup > Device Date/Time… to set the internal clock
of the device to the correct date and time.
4. Perform a manual capacitor formation.
a. Dump any charge on the capacitors.
b. Perform a test charge to full energy.
c. Retrieve the charge data.
d. Do not dump the stored charge. Allow the stored charge to dissipate for at least
10 min; the dissipation forms the capacitors.
e. If the reported charge time is clinically unacceptable, contact a Medtronic
representative.
5. Program the therapy and pacing parameters to values appropriate for the patient.
Ensure that tachyarrhythmia detection is not programmed to On.
Notes:
●
Do not enable a pacing feature that affects the pacing rate (for example, Ventricular
Rate Stabilization) before implanting the device. Doing so may result in an elevated
pacing rate that is faster than expected.
●
Patient information typically is entered at the time of initial implant, and it can be
revised at any time.
5.2 Selecting and implanting the leads
Use the guidelines in this section to select leads that are compatible with the device. The
appropriate techniques for implanting the leads may vary according to physician preference
and the patient’s anatomy or physical condition. Consult the technical manuals supplied
with the leads for specific implant instructions.