![Medtronic PROTECTA XT DR D314DRG Скачать руководство пользователя страница 104](http://html.mh-extra.com/html/medtronic/protecta-xt-dr-d314drg/protecta-xt-dr-d314drg_clinician-manual_1763148104.webp)
1. After starting a patient session, select Reports > Preferences….
2. From the Index selection box, select the Initial Report option.
3. Select the check box next to “Print Initial Interrogation Report after first interrogation”,
if desired. The report prints automatically at the beginning of a patient session after the
device is interrogated.
4. Select the additional reports to include in the Initial Interrogation Report.
5. Select [OK].
6. To print an Initial Interrogation Report for a patient session that is in progress, end and
restart the patient session. The Initial Interrogation Report prints automatically after
interrogation.
Initial Report preferences take effect at the start of a new session and remain in effect until
you change them and start a new session.
4.15.1.3 How to set Final Report preferences
You can select the reports you want printed as a part of the Final Report. The Session
Summary Report always prints when a Final Report print request is made.
1. Before ending a patient session, select Reports > Preferences….
2. From the Index selection box, select the Final Report option.
3. The Session Summary check box is selected and cannot be unselected. This ensures
that at least one report prints when a Final Report print request is made.
4. If this is the first time you are establishing Final Report preferences, the Parameters –
All Settings selection should be selected.