Q-Sense Technical Reference Manual
Proprietary
150 of 188
Changes, additions or maintenance to the equipment carried out
by persons without appropriate qualifications and training and /or
using un approved spare parts may lead to serious risk of injury
and damage to the equipment as well as making the warranty void.
17.2
Periodical Thermode Calibration
A periodical Thermode Calibration is necessary for the Q-SENSE system performance.
Perform Thermode Calibration every six months, according to the Thermode calibration service
manual.
17.3
Periodical Thermode Maintenance
1.
Inspect Thermode to confirm mechanical integrity and no visual damage before use on patient.
2.
Replacing Thermode once a year is needed to keep System integrity. Please contact Medoc
representative in your Area for more information on Thermode replacement and warranty
extended plan.