Medline ReadyBath MSCWARMER 14S Скачать руководство пользователя страница 1

Hit 

ONCE

Use arrows to

increase or decrease

temperature

RECOMMONDED

 

FACTORY 

DEFAULT

125°F

TO ENTER SET-POINT TEMPERATURE

See 

Unit

PAGE 1

     

 

 

Power Source

:   

      

AC 115 (+/-10%), 60HZ

 Max Watts: 300 (14S and 28S)

Maximum Temperature:    

54oC / 130oF

Operating Environment:

   

Indoor use, altitude up to 6,500 ft (2,000m). 

 

 

 

 

Installation Category II, Pollution Degree 2, 

 

 

 

     

ambient Conditions 10oC / 50oF to 35oC /95oF, 80% RH maximum.

Storage Temperature:      

 

-10oC / 14oF to 70oC / 158oF, 70% RH maximum

Approvals:                         

Underwriter's  UL / CUL,  United States / Canadian for laboratory equipment.

Compliance:                      

UL Standard  61010-1,   IEC 61010-1 .

 

   

Exterior

:  Powder Coated Steel   

 

 

Interior:  

Aluminum 

   

 

Insulation:

 

 

Fiberglass 

    

Door:  

Acrylic 

    

Thermo-control: 

 PID Microprocessor   

 

Heater: 

Resistive-Tubular Incoloy 

Unit Construction

MHSe WARMER14S   

  

 

 

MHSe WARMER28S

Capacity:  

14 

ReadyBaths 

   

Capacity:  

28 ReadyBaths

Exterior Dimensions: 

12.87"(32.6 cm)W  x 

 

Exterior Dimensions: 

22.14" (56.2cm)W x  

18.00" (45.7 cm)H x 13.43" (34.1 cm)D 

  

 

18.00" (45.7 cm)H x 13.25" (33.6cm)D 

Shipping Weight:  

29 pounds / 13.1 kg 

 

 

Shipping Weight:  

34 pounds / 15.4 kg

Net Weight:  

23.5

 

pounds / 10.6 kg 

 

 

Net Weight:  

28.5 pounds / 12.9 kg

Unit Specifications

The control's self diagnostics are part of Intelligent Warmer's "watch-dog" system that continuously monitor the warmer's 

operating system, temperature and performance. In the event of any of the operating alarm conditions, an alarm prompt 

will continuously appear on the control display. The alarm conditions are described below. All alarm conditions will 

automatically discontinue heater operation.  

Medline Industries, Inc.      Three Lakes Drive      Northfield  IL  60093     1-800-MEDLINE

Control prompts will only display when a fault or alarm condition exists.

   Control Self Diagnostics

Indicates Connection Error

ALARM Codes "Err1" & "Err6" 

or

Check the wiring connections

Indicates Open Input / No Sensor Error

ALARM Code  "- - - -"  

Check connection between Controller 

and Thermocouple

Indicates Internal Memory Error

ALARM Code  "E111"  

Turn Controller OFF and On.  If problem 

remains, replace Controller

Indicates Watchdog Error

ALARM Code "Err4" 

Controller needs to be replaced

Indicates Thermocouple Sensor Error

ALARM Code  "Err8"  

Turn off and wait until ambient 

temperature recovers

Indicates Sensor Temp Over/Under Error

ALARM Code  "nnnn" & "vvvv"  

Increase or reduce process temperature

to get back in range

or

Operating Specifications

 

Issue Rev C 10 19

Controller Functions Buttons

MENU

Functions

Decrease

Increase











The warmers’ controller allows you to set and change the 

units operating temperature between 100°F and 130°F, 

and change the read out units to either °C or °F. These are 

the only settings that can be changed. There are other 

controller settings that are visible, but for safety and opera-

tional purposes can not be changed.

Hit 

TWICE

Use arrows to

change temperature

units readout

FACTORY 

DEFAULT

°F

TO CHANGE TEMPERATURE READ OUT

See 

Unit

For questions regarding:

 Technical  Support  

General Product Information, please contact:       

Technical Support/Parts Order, please contact:      PH:  1-800-482-HEAT      FAX 773-622-2282

Copyright Quincy Lab Inc. 2018

Warmer Maintenance  

Periodic cleaning is the only required maintenance for the Warmer.        Disconnect the power cord before cleaning.  To 

clean interior and exterior surfaces, use a soft damp cloth.  Avoid excessive amounts of water that could drip into the lower 

plenum.  Use a mild detergent or cleaner if necessary.   
The door should only be cleaned using a lint-free cloth.  Paper towels can mar the surface of the door.  Do Not use any 

commercial or alcohol-based cleansers on the door.  This will cause cracking of the acrylic over time.

Medline Industries, Inc.  Three Lakes Drive, 

Northfield, IL 60093 

1-800-MEDLINE (633-5463) 

medline.com | [email protected]

PAGE 4

To return to temperature display, after making any change, simply press the           button until you see the warmer’s temperature 

on the display. If you are in an area of the MENU not shown press and hold           until you see the warmer’s current temperature.

P

P

Models MSCWARMER 14S & 28S

Operating Manual

NEXTGEN Warmer

INTELLIGENT W

ARMER

125

P

F

WARMING - Solid Red LED - Pack has not yet reached required temperature
HEATED - Solid Green LED - Packs have reached the required temperature and are ready to use.
TAKE FIRST - Green LED Flashes - Pack has been in the warmer for 84 hours. These packs should be used first.
DISCARD - Red LED Flashes - Pack has been in the warmer longer than 250 hours and should be discared.

INTELLIGENT W

ARMER

125

P

F

WARMING - Solid Red LED 

- Pack has not yet reach

ed required temperature

HEATED - Solid Green LED -

 Packs have reached the required

 temperature and are rea

dy to use.

TAKE FIRST - Green LE

D Flashes - Pack has bee

n in the warmer for 84 hou

rs. These packs should

 be used first.

DISCARD - Red LED Flas

hes - Pack has been in t

he warmer longer than 25

0 hours and should be discared.

Model 28S

Model 14S