125°F
TO ENTER SET-POINT TEMPERATURE
See
Unit
PAGE 1
Power Source
:
AC 115 (+/-10%), 60HZ
,
Max Watts: 300 (14S and 28S)
Maximum Temperature:
54oC / 130oF
Operating Environment:
Indoor use, altitude up to 6,500 ft (2,000m).
Installation Category II, Pollution Degree 2,
ambient Conditions 10oC / 50oF to 35oC /95oF, 80% RH maximum.
Storage Temperature:
-10oC / 14oF to 70oC / 158oF, 70% RH maximum
Approvals:
Underwriter's UL / CUL, United States / Canadian for laboratory equipment.
Compliance:
UL Standard 61010-1, IEC 61010-1 .
Exterior
: Powder Coated Steel
Interior:
Aluminum
Insulation:
Fiberglass
Door:
Acrylic
Thermo-control:
PID Microprocessor
Heater:
Resistive-Tubular Incoloy
Unit Construction
MHSe WARMER14S
MHSe WARMER28S
Capacity:
14
ReadyBaths
Capacity:
28 ReadyBaths
Exterior Dimensions:
12.87"(32.6 cm)W x
Exterior Dimensions:
22.14" (56.2cm)W x
18.00" (45.7 cm)H x 13.43" (34.1 cm)D
18.00" (45.7 cm)H x 13.25" (33.6cm)D
Shipping Weight:
29 pounds / 13.1 kg
Shipping Weight:
34 pounds / 15.4 kg
Net Weight:
23.5
pounds / 10.6 kg
Net Weight:
28.5 pounds / 12.9 kg
Unit Specifications
The control's self diagnostics are part of Intelligent Warmer's "watch-dog" system that continuously monitor the warmer's
operating system, temperature and performance. In the event of any of the operating alarm conditions, an alarm prompt
will continuously appear on the control display. The alarm conditions are described below. All alarm conditions will
automatically discontinue heater operation.
Medline Industries, Inc. Three Lakes Drive Northfield IL 60093 1-800-MEDLINE
Control prompts will only display when a fault or alarm condition exists.
Control Self Diagnostics
Indicates Connection Error
ALARM Codes "Err1" & "Err6"
or
Check the wiring connections
Indicates Open Input / No Sensor Error
ALARM Code "- - - -"
Check connection between Controller
and Thermocouple
Indicates Internal Memory Error
ALARM Code "E111"
Turn Controller OFF and On. If problem
remains, replace Controller
Indicates Watchdog Error
ALARM Code "Err4"
Controller needs to be replaced
Indicates Thermocouple Sensor Error
ALARM Code "Err8"
Turn off and wait until ambient
temperature recovers
Indicates Sensor Temp Over/Under Error
ALARM Code "nnnn" & "vvvv"
Increase or reduce process temperature
to get back in range
or
Operating Specifications
Issue Rev C 10 19
Controller Functions Buttons
MENU
Functions
Decrease
Increase
The warmers’ controller allows you to set and change the
units operating temperature between 100°F and 130°F,
and change the read out units to either °C or °F. These are
the only settings that can be changed. There are other
controller settings that are visible, but for safety and opera-
tional purposes can not be changed.
Hit
TWICE
Use arrows to
change temperature
units readout
FACTORY
DEFAULT
°F
TO CHANGE TEMPERATURE READ OUT
See
Unit
For questions regarding:
Technical Support
General Product Information, please contact:
Technical Support/Parts Order, please contact: PH: 1-800-482-HEAT FAX 773-622-2282
Copyright Quincy Lab Inc. 2018
Warmer Maintenance
Periodic cleaning is the only required maintenance for the Warmer. Disconnect the power cord before cleaning. To
clean interior and exterior surfaces, use a soft damp cloth. Avoid excessive amounts of water that could drip into the lower
plenum. Use a mild detergent or cleaner if necessary.
The door should only be cleaned using a lint-free cloth. Paper towels can mar the surface of the door. Do Not use any
commercial or alcohol-based cleansers on the door. This will cause cracking of the acrylic over time.
Medline Industries, Inc. Three Lakes Drive,
Northfield, IL 60093
1-800-MEDLINE (633-5463)
medline.com | [email protected]
PAGE 4
To return to temperature display, after making any change, simply press the button until you see the warmer’s temperature
on the display. If you are in an area of the MENU not shown press and hold until you see the warmer’s current temperature.
P
P
Models MSCWARMER 14S & 28S
Operating Manual
NEXTGEN Warmer
INTELLIGENT W
ARMER
125
P
F
WARMING - Solid Red LED - Pack has not yet reached required temperature
HEATED - Solid Green LED - Packs have reached the required temperature and are ready to use.
TAKE FIRST - Green LED Flashes - Pack has been in the warmer for 84 hours. These packs should be used first.
DISCARD - Red LED Flashes - Pack has been in the warmer longer than 250 hours and should be discared.
INTELLIGENT W
ARMER
125
P
F
WARMING - Solid Red LED
- Pack has not yet reach
ed required temperature
HEATED - Solid Green LED -
Packs have reached the required
temperature and are rea
dy to use.
TAKE FIRST - Green LE
D Flashes - Pack has bee
n in the warmer for 84 hou
rs. These packs should
be used first.
DISCARD - Red LED Flas
hes - Pack has been in t
he warmer longer than 25
0 hours and should be discared.
Model 28S
Model 14S