PULSE OXIMETER

GENERAL DESCRIPTION

PRECAUTIONS FOR USE

CAUTION

INTENDED USE

PRODUCT SPECIFICATIONS

OPERATION INSTRUCTIONS

The measurement of oxygen saturation of arterial blood (also known as pulse 

oxygen saturation, usually shortened as SpO�) adopts the principles of light 

spectra and volume tracing. The LED emits lights with two specific wavelengths, 

which are selectively absorbed by oxygenated hemoglobin and 

deoxyhemoglobin. The optical receptor measures the changes in the light 

intensity after the light passes the capillary network and estimates the ratio of 

oxygenated hemoglobin and the total hemoglobin.

SpO�% = 

x 100%

oxygenated hemoglobin

oxyhemo deoxyhemoglobin

■ 

 

  Explosion hazard. Do not use the oximeter in the presence of  flammable

      anesthetics mixture with air, oxygen, or hydrogen.

   When the oximeter is in use, there should not be any great power appliances

      as high voltage cables, X-ray machine, ultrasound equipment and electrizer in

      use nearby. 

   Keep the oximeter away from dust, vibration, corrosive substances, explosive 

      materials, high temperature and moisture.

    This oximeter does not have alarm function; please do not use this product in 

      the environment where alarm is required.

   The oximeter should be handled with care so as to avoid shocks and falls. 

    When the oximeter is in use, it must be ensured the batteries have sufficient 

      capacity; otherwise there might be such phenomena as hstarting-up 

      abnormalities or inaccurate measurement data, etc.   

   Please do not use such pointed objects as pen point or nails for pressing 

      operation, otherwise it might cause permanent damage to the surface of the

      keyboard.

   Do not make any clinical judgments based solely on the oximeter. The 

      oximeter is intended only as an adjunct in patient assessment. It must be used

      in conjunction with clinical signs and symptoms, as well as doctor’s diagnoses.

   To ensure accurate performance and prevent device failure, do not expose the

      oximeter to extreme moisture, such as direct exposure to rain. Such exposure

      may cause inaccurate performance or device failure.

   Do not conduct SpO� measurement on the finger smeared with nail polish, 

      otherwise this will lead to unreliable measurement results.

    Please do not open the enclosure. The enclosure shall only be opened by the

      authorized person.

   In order to have more accurate measurements of SpO� and PR, the oximeter

      should be used in quiet and comfortable environment.

    Follow local ordinances and recycling instructions regarding disposal or 

      recycling of the device and device components, including batteries.

    Prolonged continuous monitoring may increase the risk of unexpected 

      changes in skin characteristics, such as irritation, reddening, blistering or 

      burns. Inspect the sensor site every two hours and move the sensor if the skin

      quality changes. 

   Pulse oximeter simulator can not be used to access the accuracy of the pulse 

      oximeter. 

   The expected service life of the device is three years.

The Fingertip Pulse Oximeter is intended to measure functional arterial oxygen 

saturation (SpO�) and pulse rate of adult, pediatric and adolescent patients in 

hospital, hospital type facilities, and in the home care envirorment.

The oximeter is not suitable to monitor patients continuously for long term.

LANYARD INSTALLATIONS

BATTERY INSTALLATIONS

NOTES: Be sure to install batteries as 

directed by the polarity indicators in the 

battery cabin. Batteries in this device are 

different than conventional battery 

installation. 

   Please put or remove batteries in right

      order, or it is likely to damage the device  

      bracket.

   Battery polarities must be correctly 

      installed. Otherwise, damage might be 

      caused to the device. 

1.   Press the button down on the back panel of the oximeter.

2    Push the battery cover horizontally along the arrow shown on the back panel 

      of the oximeter.

3.   Install two AAA batteries into the battery cabin.

4.  Close the battery cover. 

MAINTENANCE

1.   Use a soft cloth dampened with either a commercial, nonabrasive cleaner, or a

      solution of 70% alcohol in water, lightly wipe the surfaces of the oximeter.

2.   Avoid using alcohol-based, amido or acetone-based detergents.

3.   The casing of the oximeter should be kept from the contamination of filth and

      dirt, and it can be wiped with non-velvet soft cloth. When cleaning, do not 

      spill the liquid onto the instrument. Ensure no liquid is allowed to enter the 

      inside of the oximeter.

4.   It is forbidden to use such grinding materials as wire brush or metal polishing 

      agent, because these materials may cause damage to the panels of the 

      oximeter. 

5.  Please do not soak the oximeter in liquid.

6.   Under normal circumstances, it is unnecessary for the oximeter to have 

      special maintenance, and cautions must be exercised on the following points 

      during the use of the oximeter: 

   Please use the oximeter in the environment according to the requirements of

      the performance criteria.

    Avoid exposure or direct sunlight. 

    Avoid excessive radioactive infrared rays or ultraviolet rays. 

    Avoid contacts with organic solutions, dusts or corrosive gases.

Measurement specifications

SpO�

Measuring Range

0~100%

Resolution

1%

Accuracy

At 70~100%, ±2%;  At 0~69%, unspecified 

Data updated period

<13 s

Accuracy in the discrete 

SpO� ranges

At 70%~80%, ±2%;  At 80%~90%, ±2%; 

At 90%~100%, ±2%

PR

Measuring Range

25~250 bpm

Resolution

1 bpm

Accuracy

±1% or ± 1 bpm, whichever is greater

Data updated period

<13 s

Battery specifications

Type

Voltage

two AAA alkaline battery

1.5 Volts DC (per battery)

The oximeter uses two 1.5 V AAA type batteries and a set of new batteries can be 

used for more than 18 hours, depending on  concrete battery types.

1.    Install batteries as indicated by the battery 

      installation instructions. 

2.  Open the oximeter. Insert one finger into the 

      rubber hole of the oximeter before releasing. 

       Your nail surface must be upward. 

3.  Press the function button once on the front panel.

4.  Read corresponding data on the display screen. 

5.   After turning on the oximeter, each time you press the power switch, the

      display screen will change to another direction.  If you hold down the power 

      switch, you can adjust the brightness of the screen from 1 to 5. 

1.    Attach the lanyard through the hole in the middle of the front face as the  

      arrows show. 

2.   Complete.

The mechanical activity of the heart causes arterial pulse, by measuring the 

pulse we can get PR value. 

The oximeter is standalone, reusable, and not reprocessed. The sensor of the 

oximeter is built-in.

■ 

 

  Federal Law Restricts this device to sale by or on the order of a physician. 

   Please read the user manual carefully prior to operating. 

  SpO� Plethysmogram (normalized)

 

SpO� reading

 Pulse rate reading

 Indication of battery capacity

 

Indication of pulse intensity

      Indication of screen brightness (Default value of brightness is 3)

      Display modes