92
10-25.
Near Addition Test using Cress Grid (Near ADD)
-
Purpose: Correct Near Addition according to Accommodation, using
Cross Grid.
-
Chart: Near Vision Chart (Cross Grid)
-
Auxiliary Lens : None
-
Objective: Make the clarity and thickness of horizontal/vertical lines
equally, similar with Far Vision Cross Grid Test.
-
Operation Sequence
1.
Press [ADD #] button
to drop the Near Vision Cross
Grid chart at the place of about 40cm.
2.
Press R/B/L button to select the eye for test.
3.
Adjust the ADD value until the thickness or clarity of
horizontal/vertical line becomes equal.
-
When the vertical line is seen clearly, turn the dial clockwise
(- direction)
-
When the horizontal line is seen clearly, turn the dial
counterclockwise (+ direction.)