2-8-338
HOOKUP CONNECTION GUIDE
Using MEDIVATORS
®
ADVANTAGE
®
Automated Endoscope Reprocessors
ADVANTAGE HOOKUP DIAGRAM
Representative endoscope for illustrative purposes only. For exact
port locations, refer to endoscope manufacturer’s instruction guide.
Each ADVANTAGE Hookup is comprised of multiple ports and connector assemblies.
Each port is labeled with a Port ID number (circled). This same number also identifies the
corresponding Connector Assembly.
READ ALL INSTRUCTIONS PRIOR TO CONNECTION
Locate the endoscope port and follow the instructions for attaching the
corresponding Endoscope Connector. All connectors must be attached.
LOT XXXXXX ID: XXXX
P/N: X-X-XXX
Medivators Inc.
14605 28th Ave N.
Minneapolis, MN 55447
USA
Medivators BV
Sourethweg 11
6422PC Heerlen
The Netherlands
DESIGNATED HOOKUP FOR THE FOLLOWING FUJIFILM™ ENDOSCOPES*:
EB-530US
* This list of corresponding endoscope models is subject to change without notice. For a full list, please go to:
www.medivators.com/hookuplookup
© 2015 Medivators Inc. 50098-1019-EN/A
Port ID
Port Name
Connector Assembly Description
Order Number
LEAK TEST
Leak Test Connector Assembly
H28338-1
JET/AUX
Air/Water Connector Assembly
N/A
BIOPSY
Biopsy Inlet Connector Assembly
H28338-3
AIR
Air Probe Connector Assembly
N/A
ELEVATOR
Auxiliary Water Connector Assembly
N/A
WATER
Air/Water Connector Assembly
N/A
SUCTION
Suction Connector Assembly
H28338-7
EXTRA
Balloon Connector Assembly
H28338-8
If a Connector Assembly needs replacement, locate the port number/name and use
the corresponding order number found in the chart above. Connector Assemblies are
specific to individual hookups.
If you have any questions regarding this hookup, in the U.S., please contact Medivators
Technical Support at 1-800-444-4729. Outside the U.S., please contact your Medivators
Distributor.
CONNECTOR ASSEMBLY IDENTIFICATION CHART
Leak Test Connector
Align pins on connector with slots
on the scope. Push on and turn
clockwise to engage.
Suction Connector
Align stainless steel fitting
with Fujinon Cleaning
Adaptor, push and turn
clockwise to connect.
Balloon Connector
Align stainless steel fitting with
balloon port, push and turn
clockwise to connect.
Biopsy Inlet Connector
Align stainless steel fitting with
Fujinon Cleaning Adaptor, push
and turn clockwise to connect.
FUJIFILM Cleaning
Adaptor WA-003
Refer to FUJIFILM User
Manual for connection
instructions
FUJIFILM
Cleaning Adaptor
WA-001
Refer to FUJIFILM User
Manual for connection
instructions
Ensure Biopsy Connections are correct to avoid alarms.