2-8-611HAN/2-8-611CAS 

HOOKUP CONNECTION GUIDE 

Using MEDIVATORS

®

 ADVANTAGE

®

 Automated Endoscope Reprocessors

Port ID

Port Name

Connector Assembly Description

Order Number

1

LEAK TEST

Leak Test Connector Assembly

H28611-1

2

JET/AUX

Air/Water Connector Assembly

H28611-26

3

BIOPSY

Biopsy Connector Assembly

H28611-3

4

AIR

Air Probe Connector Assembly

H28611-4

5

ELEVATOR

Elevator Wire/Auxiliary Water Connector Assembly

H28611-5

6

WATER

Air/Water Connector Assembly

H28611-26

7

SUCTION

Suction Connector Assembly

H28611-7

A

NA

Air/Water & Suction Port Separator

1-5-112

If a Connector Assembly needs replacement, locate the port number/name and use 
the corresponding order number found in the Connector Assembly Identification Chart. 
Connector Assemblies are specific to individual hookups. Check each Hookup Connection 
Guide for relevant information.

ADVANTAGE HOOKUP and CHANNEL SEPARATOR  

(Figure 1)

* Representative endoscope for illustrative purposes only. For exact 

port locations, refer to endoscope manufacturer’s instruction guide.

Each Advantage Hookup is comprised of multiple ports and connector assemblies. Each 
port is labeled with a Port ID number (circled). This same number also identifies the 
corresponding Connector Assembly. Each Connector Assembly is fitted with a specific 
endoscope connector. (see Fig. 2).

EXAMPLE CONNECTOR ASSEMBLY 

(Figure 2)

READ ALL INSTRUCTIONS PRIOR TO CONNECTION

Locate the endoscope port* and follow the instructions for attaching the  

corresponding Endoscope Connector. All connectors must be attached.

CONNECTOR ASSEMBLY IDENTIFICATION CHART

Ferrule

Endoscope

Connector

example

Tubing

Connector 

Coupling

Label

Ensure proper alignment of all connectors with the 

endoscope mating parts prior to attachment.

A



Leak Test Connector

Align the slot on connector with the pin on the 

scope. Push on and turn clockwise to engage.
To remove leak test connector, turn connector 

counter-clockwise until slot is aligned with pin and 

pull off. Do not pull until slot is aligned with pin.

Suction Connector

Align silicone tubing with 

suction barb and push tubing 

onto barb until fully engaged.

Elevator Wire Channel Connector 

(endoscope model dependent)

Align the connector with the endoscope fitting 

and screw clockwise until fully engaged. (Check 

alternate location five)

Air/Water and Suction 

Port Separator

Align the separator with 

piston ports and push to 

compress springs before 

sliding the frame forward to 

fully engage.

A

Air Probe Connector

Align with Air Probe, push and 

slide until fully engaged.

Biopsy Connector

Align the biopsy 

connector disc over the 

biopsy inlet and push 

down to compress spring 

and slide the frame 

forward to fully engage.

Air/Water Connector

Align the connector 

openings with the air/water 

ports of the endoscope as 

shown. Ensure side latch 

is in. Push connector onto 

ports until latch moves 

out and clicks. To remove, 

push latch to release.



Auxiliary Water Connector 

(endoscope model dependent)

Align the connector with the endoscope 

fitting and screw clockwise until fully 

engaged. (Check alternate location five)

If you have any questions regarding this hookup, in the U.S., please contact MEDIVATORS Technical Support 

at 1-800-444-4729. Outside the U.S., please contact your MEDIVATORS Distributor.