6 Troubleshooting
6.1 Sensor out range / under range
Indication
Error messages from the software indication sensor out of range
Reason for error/fault
Mechanical damage to the sensor.
LoadCell not sending correct values to Phidgets bridge.
Corrective actions
Replace LoadCell
6.1.1 Cover removal
1.
Use Hex Key No.3
A.
Remove three covers and
place them in a secure
place
6.1.2 Back pulley release
1.
Extension bar 1/2"-10"
2.
Ratchet socket 1/2"-17mm
3.
Bit to Socket 1/2" adapter
4.
Cordless electronic drill
A.
Release two M10 bolts
from pulley.
Note:
BE CARFUL NOT DAMGE THE
INNER CABLE LOCATED IN
THE RIGHT HOLE
.
B.
Place the pulley on top of
the running board
C.
Cut black cable tie
Handling with heavy components, Side covers serviced with two people only, service
with power off and disconnection from main.