4.3.6 Limit switch safety operation
Purpose
Checking if at maximal lateral perturbation the limit switch does not activated.
Procedure
Access to Setting
Technical
Treadmill
Press 21.75, -21.75
Apply
The limit switch should not be activated, if yes adjust the limiters locations.
4.3.7 Homing process conducted correct
Purpose
Confirm perturbation plate return to physical zero of the device after Homing command.
Procedure
Access to Setting
Technical
Treadmill
Press Homing.
Confirm after Homing the perturbation plate returns to the center for the device.
If the plate do not returns to the exact center location, please consult with MediTouch.
4.3.8 Software update
Purpose
BalanceTutor latest software version is crucial for safety and correct functionality of the device.
Procedure
Schedule with MediTouch for software update.